DEPUY MITEK LLC US ANGLED SIDE EFFECT ELECTRODE, 21º; ENDOSCOPIC ELECTROSURGICAL ELECTRODE, BIPOLAR, SINGLE USE
|
Back to Search Results |
|
Model Number 225302 |
Device Problems
Use of Device Problem (1670); Blocked Connection (2888)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/28/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Additional narrative: udi: (b)(4).Reporter is a j&j employee.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|
|
Event Description
|
It was reported by the affiliate in japan that during an unknown procedure on (b)(6) 2023, it was observed that the angled side effect electrode, 21º device was not recognized although it was attached to the handpiece.Another like device was used to complete the procedure.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
|
|
Manufacturer Narrative
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will e filed as appropriate.H10 additional narrative: investigation summary: the complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.Since the complaint device was discarded, we cannot determine a root cause for the reported failure.If additional information is received in the future, we will reopen the complaint and perform the investigation as appropriate.A manufacturing record evaluation was performed for the finished device lot number (u2205023), and no non-conformances were identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
|
|
Search Alerts/Recalls
|
|
|