The subject device was returned for evaluation and the customer¿s reported problem was confirmed.The ceramic insulation at the distal tip of the device was found loose and worn.The inspection also noted the following (non-reportable) defects, the laser marking on the device is partially faded.The device has not been repaired in the past year.Since the lot number of the device is unknown or could not be confirmed, a device history records review could not be performed.The root cause of the loose and worn ceramic insulation at the distal tip of the sheath cannot conclusively be determined.However, the most likely causes are due to: thermally and/or mechanically fatigue, wear and tear, improper handling (more specifically the device being subjected to mechanical overload, impact, accidental dropping, etc.).A design change that was implemented in 2010 to prevent reoccurrence.To mitigate damage to the device and patient injury, the instruction for use provides the following: - properly reprocess the product before first and each subsequent use following the instructions in this manual and in the system guide endoscopy.Improper and/or incomplete reprocessing can cause infection of the patient and/or medical personnel.- visually inspect the product.Make sure that it has: no corrosion, no dents, no scratches.- visually inspect the ceramic insulation at the sheath¿s distal end before each use.Do not use the instrument in case of damage (e.G., cracks, fractures).- impact, fall, shock, or similar stress can damage the ceramic insulation at the sheath¿s distal end.Damaged instruments can cause injuries to the patient and/or user.Do not use the instrument if damaged.- if the product is damaged or does not function properly, contact an olympus representative or an authorized service center.In general, the end-user is required to inspect the device for defects, check the function of all devices and have alternate equipment prior to use.Olympus will continue to monitor complaints related to the device and reported phenomenon.Additional 510(k): k931995.
|