BECTON, DICKINSON AND COMPANY (BD) BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE
|
Back to Search Results |
|
Model Number 367957 |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/25/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
E.1.Initial reporter phone #: (b)(6).H.6.Investigation summary: bd had not received samples, but one (1) photo was provided for investigation.The photo was reviewed and the indicated failure mode for molding defects was observed.Additionally, one hundred (100) retention samples from bd inventory were evaluated by visual examination and the issue of molding defects was not observed.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint has been confirmed for the indicated failure mode molding defects.Bd was not able to identify a root cause for the indicated failure mode.H3: see h.10.
|
|
Event Description
|
It was reported that prior to use with bd vacutainer® sst¿ ii advance plus blood collection tubes the lid was discovered to be broken.The following information was provided by the initial reporter, translated from chinese to english: the patient suffered from pain and discomfort under the xiphoid process for 1 month, and was diagnosed with hydrocardia.At 07:36 on (b)(6) 2023, he was admitted to the hospital at 07:36.When the nurse took blood for his blood routine examination, he found that the shield of the vacuum blood collection tube was damaged and could not be used normally.He immediately replaced it with another vacuum blood collection tube for use.
|
|
Search Alerts/Recalls
|
|
|