• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON AND COMPANY (BD) BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON AND COMPANY (BD) BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Model Number 367957
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/25/2023
Event Type  malfunction  
Manufacturer Narrative
E.1.Initial reporter phone #: (b)(6).H.6.Investigation summary: bd had not received samples, but one (1) photo was provided for investigation.The photo was reviewed and the indicated failure mode for molding defects was observed.Additionally, one hundred (100) retention samples from bd inventory were evaluated by visual examination and the issue of molding defects was not observed.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint has been confirmed for the indicated failure mode molding defects.Bd was not able to identify a root cause for the indicated failure mode.H3: see h.10.
 
Event Description
It was reported that prior to use with bd vacutainer® sst¿ ii advance plus blood collection tubes the lid was discovered to be broken.The following information was provided by the initial reporter, translated from chinese to english: the patient suffered from pain and discomfort under the xiphoid process for 1 month, and was diagnosed with hydrocardia.At 07:36 on (b)(6) 2023, he was admitted to the hospital at 07:36.When the nurse took blood for his blood routine examination, he found that the shield of the vacuum blood collection tube was damaged and could not be used normally.He immediately replaced it with another vacuum blood collection tube for use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBES
Type of Device
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
UK 
Manufacturer (Section G)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
UK  
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8448235433
MDR Report Key16981894
MDR Text Key315764972
Report Number9617032-2023-00592
Device Sequence Number1
Product Code JKA
UDI-Device Identifier50382903679572
UDI-Public(01)50382903679572
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
BK050036
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2023
Device Model Number367957
Device Catalogue Number367957
Device Lot Number2084349
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/09/2023
Initial Date FDA Received05/22/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/25/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-