Catalog Number 00514010400 |
Device Problems
Thermal Decomposition of Device (1071); Circuit Failure (1089); Failure to Power Up (1476)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/01/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that outside of surgery on (b)(6) 2023 the unit would not power up as it was not getting any power.The fuses were also burnt.There was no patient involvement.No adverse events were reported as a result of this malfunction.Due diligence is complete.No further information is available.
|
|
Manufacturer Narrative
|
This event is recorded by zimmer biomet under (b)(4).The product will not be returned to zimmer biomet.Product is being evaluated by external contractor.Once the investigation is complete, a follow up/final report will be submitted.
|
|
Event Description
|
No additional event information available.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined the unit was not getting power and fuses burnt and power inlet module replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|