• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HILL-ROM BATESVILLE AFFINITY 4 BED FRAME; TABLE, OBSTETRIC (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM BATESVILLE AFFINITY 4 BED FRAME; TABLE, OBSTETRIC (AND ACCESSORIES) Back to Search Results
Model Number P3700E000005
Device Problem Activation Problem (4042)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/24/2023
Event Type  malfunction  
Manufacturer Narrative
The hillrom technician found the cpr block needed to be replaced.The affinity® 4 birthing bed requires an effective maintenance program.We recommend that you do annual preventive maintenance (pm) for joint commission certification.Pm not only meets joint commission requirements but can help make sure of a long, operative life for the affinity® 4 birthing bed.Two effective ways to reduce downtime and make sure the patient remains comfortable are to keep accurate records and maintain the affinity® 4 birthing bed.Test the cpr release for proper operation and reset of the head drive system.When the cpr release is pulled, the bed should lower from any position into a flat position within 7 seconds with at least 50 lb (23 kg) of weight on the head section.A search of the hillrom maintenance records showed hillrom performed preventative maintenance on this bed on march 2023.It is unknown if the facility performed any other preventative maintenance on this bed.The technician replaced the cpr block to resolve the reported event.Based on this information, no further action is required.
 
Event Description
Hillrom received a report from a hillrom technician stating the cpr was inoperative.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFFINITY 4 BED FRAME
Type of Device
TABLE, OBSTETRIC (AND ACCESSORIES)
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
brad wheeler
1069 state route 46 east
batesville, IN 47006
3128199307
MDR Report Key16986760
MDR Text Key315847434
Report Number1824206-2023-00443
Device Sequence Number1
Product Code KNC
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K915779
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 05/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberP3700E000005
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/24/2023
Initial Date FDA Received05/23/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/29/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-