Model Number BI70002000 |
Device Problem
Communication or Transmission Problem (2896)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/03/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: bi71000896, serial/lot #: unknown; product id: bi71000167, serial/lot #: unknown; product id: bi71000424, serial/lot #: unknown; product id: bi71000319, serial/lot #: unknown.H3: a medtronic representative went to the site to test the equipment.Testing revealed that the rear cab breakout panel and umbilical cable were replaced.The system was tested with busmaster to verify that there was no longer any noise on the can.The imaging system then passed the system checkout and was found to be fully functional.H6: fdm b01, fdr c13, fdc d02 are applicable to the system checkout.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information regarding an imaging system being used outside of a procedure.It was reported that after the 4.2.1 upgrade, the mobile viewing system (mvs) and image acquisition system (ias) had issues communicating.There was no patient involvement.Troubleshooting included using busmaster to isolate the noise on the can ultimately leading to either the umbilical cable or the rear cab breakout panel.
|
|
Manufacturer Narrative
|
Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: bi71000896, serial/lot#: (b)(6), h2) the mobile view station (mvs) computer was returned to the manufacturer for analysis.After functional and performance testing, visual/physical examination, and usability inspection, the reported issue was not confirmed.The operating system and imaging system application 4.2.0 loaded successfully.The field representative installed the mobile view station (mvs) server in the imaging system.The system initialized.The motion, generator, communication, and charging readied.Two dimensional and three dimensional images passed.Codes: b01, c19, d14.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Manufacturer Narrative
|
H3, h6: a hardware analysis was initiated for (b)(6) to determine the probable cause of the reported behavior.Analysis found that the complaint was not verified and there were no failures found.Code b01 is applicable and previously reported.Codes c19, d14 are applicable.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Manufacturer Narrative
|
The (b)(4) lot: 71750 rev.E was returned to the manufacturer for analysis.After functional testing and visual and physical examination, the reported complaint was not confirmed.Visual inspection showed normal wear with all connections passing continuity testing and were good and secure.No faults were found.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|