• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION MOMENTUM X4 CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION MOMENTUM X4 CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D Back to Search Results
Model Number G138
Device Problem Failure to Capture (1081)
Patient Problems Undesired Nerve Stimulation (1980); Discomfort (2330)
Event Date 05/08/2023
Event Type  Injury  
Manufacturer Narrative
If pertinent information is provided in the future, a supplemental report will be submitted.
 
Event Description
It was reported that this device exhibited left ventricular (lv) loss of capture and diaphragmatic stimulation in all vectors.The physician opted to implant a device along with a new left bundle lead.This device was explanted.No additional adverse patient effects were reported.
 
Event Description
It was reported that this device exhibited left ventricular (lv) loss of capture and diaphragmatic stimulation in all vectors.The physician opted to implant a device along with a new left bundle lead.This device was explanted.No additional adverse patient effects were reported.
 
Manufacturer Narrative
If pertinent information is provided the returned device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.The device was then exposed to simulated heart load conditions, and the defibrillation, pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOMENTUM X4 CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key16988626
MDR Text Key315818941
Report Number2124215-2023-25728
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526589188
UDI-Public00802526589188
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S436
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/24/2024
Device Model NumberG138
Device Catalogue NumberG138
Device Lot Number389805
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/08/2023
Initial Date FDA Received05/23/2023
Supplement Dates Manufacturer Received06/28/2023
Supplement Dates FDA Received06/30/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/03/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age74 YR
Patient SexFemale
-
-