• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HORIZON TI MED 6/CART 180/BOX; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HORIZON TI MED 6/CART 180/BOX; CLIP, IMPLANTABLE Back to Search Results
Model Number IPN914925
Device Problems Failure to Form Staple (2579); Lack of Effect (4065)
Patient Problem Exsanguination (1841)
Event Date 05/15/2023
Event Type  Injury  
Manufacturer Narrative
(b)(4).N/a.Other remarks: n/a.Corrected data: n/a.
 
Event Description
It was reported "total thyroidectomy completed on patient.Patient transferred over to stretcher.Drain noted to be full, and patient's neck noted to be swelling.Patient transferred back onto or bed and surgical incision reopened.Surgeon noted that titanium clip had come off of vessel and was bleeding.Upon further examination surgeon noted that several (7) of the clips had come off".As a result, a procedure to properly close the surgical incision was performed.The patient status is reported as "stable".Additional information requested regarding total cartridges used.At time of this report, no additional information has been received by complainant.
 
Manufacturer Narrative
Qn# (b)(4).Associated to mdr 3003898360-2023-00811 for a different lot number used.Complaint verification testing could not be performed as no sample was returned for analysis.Part number 002200 is not being manufactured currently, however, another part number from the same family was use for the "verification of failure mode reported in the current manufacturing process" and was conducted as follows: 59 samples were taken from the current production, the samples were functionally inspected, and during the test issue reported "ligation failure-clip reopened" was not observed in the current manufacturing process.A device history record review was performed, and no relevant findings were identified.Without the device to evaluate, the complaint could not be confirmed, and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.Other remarks: n/a.Corrected data: section e.Included reporter information that was initially not reported.
 
Event Description
It was reported "total thyroidectomy completed on patient.Patient transferred over to stretcher.Drain noted to be full, and patient's neck noted to be swelling.Patient transferred back onto or bed and surgical incision reopened.Surgeon noted that titanium clip had come off of vessel and was bleeding.Upon further examination surgeon noted that several (7) of the clips had come off".As a result, a procedure to properly close the surgical incision was performed.The patient status is reported as "stable".Additional information requested regarding total cartridges used.At time of this report, no additional information has been received by complainant.
 
Manufacturer Narrative
(b)(4).Additional information received 30-may-2023 from the ent physician and the rn care facilitator at the user facility.It was confirmed that 7 clips came off the vessel.It was reported that "pt had bleeding in the mecknupon extubation that was identified immediately.He was put back inder ga.Wound opened, clit evacuated, site if bleeding identified as superior thyroid artery (clips had been sued at this site during the surgery), bleeding controlled, wound closed.Pt extubated and brought to recovery room in good condition.No long term problems." it was also reported that "pt required the surgery to evacuate the hematoma and control the bleeding.Los was not affected (he was already being admitted sda)" other remarks: n/a.
 
Event Description
It was reported "total thyroidectomy completed on patient.Patient transferred over to stretcher.Drain noted to be full, and patient's neck noted to be swelling.Patient transferred back onto or bed and surgical incision reopened.Surgeon noted that titanium clip had come off of vessel and was bleeding.Upon further examination surgeon noted that several (7) of the clips had come off".As a result, a procedure to properly close the surgical incision was performed.The patient status is reported as "stable".Additional information requested regarding total cartridges used.Associated to mdr 3003898360-2023-00811 for a different lot number used.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HORIZON TI MED 6/CART 180/BOX
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
rancho el descanso
tecate 21478
MX   21478
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key16988791
MDR Text Key315835952
Report Number3003898360-2023-00820
Device Sequence Number1
Product Code FZP
UDI-Device Identifier24026704696363
UDI-Public24026704696363
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K132658
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 05/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberIPN914925
Device Catalogue Number002200
Device Lot Number73C2300444
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/19/2023
Initial Date FDA Received05/23/2023
Supplement Dates Manufacturer Received05/19/2023
05/19/2023
Supplement Dates FDA Received05/23/2023
06/27/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/14/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-