Device analysis: returned product consisted of an innova self-expanding stent system.The outer sheath, tip, inner sheath, and the remainder of the device were checked for damage.Visual examination revealed a kink to the outer sheath at the nosecone and a kink to the inner liner, 16.2cm from the tip.The nosecone was damaged, and a portion was broken off and missing.The nosecone portion of the handle was separated.Microscopic examination revealed no additional damages.The stent appeared to have been deployed and was not returned for analysis.Because there was no evidence of any product quality deficiencies, it was considered likely that the kinked inner liner was attributable to handling after the procedure.Inspection of the remainder of the device revealed no other damage or irregularities.Product analysis found damage to the device that could have contributed to the reported event.
|