The customer, a syncardia certified hospital, reported that while troubleshooting an anomalous noise observed by one of the clinicians during patient rounds, the spring in the cpc connector on the driveline became dislodged during driver exchange.The driver was not alarming or displaying any visual indicators of an issue with the driver¿s performance.The customer determined there was no fault with the driver and retained it for use at the hospital.The cpc connector on the driveline was replaced.There was no reported impact to the patient.The cpc connector is being returned to syncardia for evaluation.
|