Patient data file review revealed three failed system checks at customer site, confirming the reported issue.All three of these tests failed the "max" left vacuum tests, where the left maximum vacuum was greater than the specification requirement (-30 to -45 mmhg); actual values were -46 mmhg and -47 mmhg.Visual inspection of internal components identified capacitor damage on the power management board.Capacitor damage is unrelated to the system check failure.Visual inspection of external components revealed no abnormalities.During the investigation, the companion 2 driver passed the system check, indicating vacuums were functioning as intended.But, as a precaution, both vacuums were replaced.The companion 2 driver failed functional testing where left drive pressure was below specification requirements.The root cause of the failure was the electronic pressure regulator.The electronic pressure regulator was rejected and replaced.Functional testing was then repeated with passing results.Failure investigation for this complaint confirmed the system check failure via patient data file review, which showed three system check failures recorded in the driver.The system check failure was not replicated via functional/observational testing.The root cause of the system check failure is vacuum pressure exceeding acceptable range allowed during system check.The root cause of the device malfunction identified during functional testing was the electronic pressure regulator, which caused the left drive pressure to not meet specification requirements.This issue will be monitored and trended as part of the customer complaint process.Syncardia has completed its evaluation and is closing this file.(if new or additional information is received in the future, syncardia will file a follow-up mdr.) (b)(4) follow-up report 1.
|