BIOSENSE WEBSTER INC THERMOCOOL® SMART TOUCH¿ ELECTROPHYSIOLOGY CATHETER; CARDIAC ABLATION PERCUTANEOUS CATHETER Back to Search Results
Model Number D133603
Device Problem Patient Device Interaction Problem (4001)
Patient Problem Stroke/CVA (1770)
Event Date 04/26/2023
Event Type  Injury  
Manufacturer Narrative
Additional information received indicates that the device is not available for return, therefore no product investigation can be performed, and the customer complaint cannot be confirmed.Manufacturing record evaluation (mre) cannot be conducted because no lot number was provided by the customer.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: (b)(4).
 
Event Description
It was reported that a patient underwent atrial flutter left (l-afl) ablation procedure with a thermocool® smart touch¿ electrophysiology catheter.The patient developed a small stroke.It was reported that 2 hours post left sided atypical atrial flutter procedure performed yesterday the patient developed a small stroke.The physician feels that it was due to the length of the procedure.The patient was stable and will be discharged home today.Additional information was received.It was discovered post use of biosense webster products.There is no information about if any intervention was performed.The physician¿s opinion on the cause of this adverse event was length of procedure, no product malfunction and procedure was a left sided flutter.Patient made a recovery and went home (b)(6) 2023.Was told by the physician the patient was ok to leave.Unsure if there are residual effects.Patient did not require extended hospitalization because of the adverse event as patient intended on an overnight stay and left the following day.Other relevant history-had previous ablation.There was no evidence of char during the procedure.There was no evidence of blood thrombus/clot during procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOCOOL® SMART TOUCH¿ ELECTROPHYSIOLOGY CATHETER
Type of Device
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key16998306
MDR Text Key315912011
Report Number2029046-2023-01144
Device Sequence Number1
Product Code LPB
UDI-Device Identifier10846835009019
UDI-Public10846835009019
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030031/S053
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 05/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberD133603
Device Catalogue NumberD133603
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/26/2023
Initial Date FDA Received05/24/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SMARTABLATE GENERATOR KIT-US
Patient Outcome(s) Life Threatening;
Patient SexMale
-
-