MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR ESSENTI
|
Back to Search Results |
|
Model Number 37612 |
Device Problems
Display or Visual Feedback Problem (1184); Failure to Interrogate (1332); Improper or Incorrect Procedure or Method (2017); Battery Problem (2885); Charging Problem (2892); Communication or Transmission Problem (2896); Insufficient Information (3190)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Event Description
|
It was reported that the patient has not been able to recharge, patient can't remember when she successfully recharged last.Patient programmer won't connect to the implantable neurostimulator (ins) neurostimulator too.Technical services(ts) tried to use the recharger to check when patient last recharged, but the device gave a 2000 date.Ts told the manufacturer's representative (rep) he'll need to meet with patient to do physician recharge mode.The issue was not resolved through troubleshooting.The rep reports that the ins is in over-discharge.The rep is trying passive mode recharge (prm) the rep wondered when they should stop prm to clear the por.Ts reviewed information about charging.The rep was able to start a normal charging session.Prior to calling, the rep tried to interrogate the ins with the clinician programmer and it showed a low battery message.No symptoms were reported.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information received from the manufacturer¿s representative (rep) reported the cause of the por/overdischarge was due to the patient not charging the implant for months, maybe even a year.A physician mode recharge was performed which resolved the por.
|
|