Model Number HL 20
Device Problem Pumping Problem (3016)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
It was reported that the hl20 pump displayed the error message: "head error" during routine check.No harm to any person has been reported.A getinge field service technician (fst) was onsite for an investigation.The fst recalibrated the optical tacho board and cleaned the tacho strobe foil.After that the error message: "head error" was not present anymore.Full functional check has been performed.The device is back in clinical use.Thus the reported failure could be confirmed.The most probable root cause could be determined as a dirty tacho strobe foil.Because the sensor no longer detects one or more marks on the tacho strobe foil, it measures an apparently low speed at the pump head.Furthermore the failure mode "head error" can be linked to the following most possible root causes according to the hl 20 risk management file: failure of pump control board.Defective/ dirty tacho, relay or pump belt.The device in question was manufactured on 2016-10-27.The review of the non-conformities during the period of 2016-10-27 to 2023-05-24 does not show any non-conformity in regard to the reported product and failure.There is no indication that manufacturing issues occurred during this time, thus production related influences are unlikely.The device caused the complaint and was not able to work as per factory¿s specifications.The occurrence rate was calculated for the reported issue and it was determined that this is not a systemic issue.Therefore, no remedial action is required.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Event Description
It was reported that the hl20 pump displayed the error message: "head error" during routine check.No harm to any person has been reported.Complaint id# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEART LUNG MACHINE
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
Manufacturer (Section G)
JULIA KAPFENBERGER
neue rottenburger strasse 37
hechingen
Manufacturer Contact
neue rottenburger strasse 37
hechingen 
MDR Report Key17001464
MDR Text Key315916302
Report Number8010762-2023-00240
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K943803
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL 20
Device Catalogue Number701028580
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/08/2023
Initial Date FDA Received05/25/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/27/2016
Is the Device Single Use? No
Type of Device Usage A
Patient Sequence Number1
-
-