Other, other text: d10.Device available for evaluation; h3.Device evaluated by manufacturer and h6.Evaluation codes: updated.One device was received.Visual inspection: cracked tank cover, faded line cord, impact damage to liquid crystal display (lcd), and corroded drain fitting.The tank cover was cracked confirming the complaint.No other analysis was performed.A root cause was due to overtightening the tank cover screws.A device history record (dhr) review was not performed because the device is beyond a year from its manufacture date and there was no indication of a manufacturing defect during the investigation.Service history review identified this device has not been in for service previously.No action taken due to the condition of the device.It is deemed beyond economical repair and will be scrapped.
|