• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OAKDALE LEVEL 1 HOTLINE FLUID WARMER, 390; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OAKDALE LEVEL 1 HOTLINE FLUID WARMER, 390; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-390
Device Problems Fracture (1260); Leak/Splash (1354)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
B3: date of event and d4: udi number is unknown.No information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported, that the fluid warmer had a cracked cover and was leaking.Restock n swap.Patient involvement unknown.
 
Manufacturer Narrative
Other, other text: d10.Device available for evaluation; h3.Device evaluated by manufacturer and h6.Evaluation codes: updated.One device was received.Visual inspection: cracked tank cover, faded line cord, impact damage to liquid crystal display (lcd), and corroded drain fitting.The tank cover was cracked confirming the complaint.No other analysis was performed.A root cause was due to overtightening the tank cover screws.A device history record (dhr) review was not performed because the device is beyond a year from its manufacture date and there was no indication of a manufacturing defect during the investigation.Service history review identified this device has not been in for service previously.No action taken due to the condition of the device.It is deemed beyond economical repair and will be scrapped.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER, 390
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17003177
MDR Text Key315939330
Report Number3012307300-2023-05656
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 06/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/25/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-390
Device Catalogue NumberCON-HL-390
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/12/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/22/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-