Catalog Number P/N 397002-001 |
Device Problems
Output Problem (3005); Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/16/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The companion 2 driver will be further investigated by syncardia.The results will be provided in a follow-up mdr.
|
|
Event Description
|
During routine evaluation, a syncardia service technician reported that the companion 2 driver has a very low cardiac output of 1.1 as well as a high reading on the left driver pressure.Driver also switches to the second compressor once the driver is turned on.
|
|
Event Description
|
During routine evaluation, a syncardia service technician reported that the companion 2 driver has a very low cardiac output of 1.1 as well as a high reading on the left driver pressure.Driver also switches to the second compressor once the driver is turned on.
|
|
Manufacturer Narrative
|
Device history record (dhr) review confirmed that companion 2 driver s/n (b)(6) was serviced and passed all functional testing prior to being released to finished goods.Alarm history and patient data file review found multiple alarms indicating "very low left pressure." visual inspection of external components found no abnormalities.Visual inspection of internal components found damage to capacitor.Driver failed functional testing for acceptance at incoming inspection due to out of range cardiac output and high left driveline pressure.Additional testing concluded the excessive left driveline pressure was due to the left driveline transducer on sensor board.A known functional sensor board was installed and driver passed all functional testing without alarm or malfunction.Failure investigation for this complaint confirmed the reported issue.The customer complaint was replicated during testing; root cause of the reported out of range outputs and system check failure was due to a faulty pressure sensor board.Failure investigation identified no other test failures or damage that could have contributed to the complaint.Damage found to the capacitor was cosmetic only.Device was not in patient use at time of complaint.This issue will be monitored and trended as part of the customer complaint process.Syncardia has completed its evaluation and is closing this file.If new or additional information is received in the future, syncardia will file a follow-up mdr.
|
|
Search Alerts/Recalls
|