• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Catalog Number 595000-001
Device Problem Audible Prompt/Feedback Problem (4020)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/28/2023
Event Type  malfunction  
Event Description
The customer, a syncardia certified hospital, reported that the driver exhibited a fault alarm while supporting a patient.There was no reported adverse patient impact.The patient was switched to a back-up driver.
 
Manufacturer Narrative
The driver will be returned to syncardia for evaluation.The results will be provided in a follow-up mdr.
 
Manufacturer Narrative
Device history record (dhr) review confirmed that freedom driver s/n (b)(6) was serviced and passed all functional testing prior to being released to finished goods.Alarm history data review and visual inspection of internal components confirmed the customer reported fault alarm condition.Visual inspection of external components found no abnormalities.The freedom driver passed all functional testing at incoming inspection and during investigation testing.Complaint could not be replicated.The failure investigation found no evidence of a device malfunction.Patient was switched to a backup driver without any reported adverse impact.This issue will be monitored and trended as part of the customer complaint process.Syncardia has completed its evaluation and is closing this file.If new or additional information is received in the future, syncardia will file a follow-up mdr.
 
Event Description
The customer, a syncardia certified hospital, reported that the driver exhibited a fault alarm while supporting a patient.There was no reported adverse patient impact.The patient was switched to a back-up driver.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e silverlake road
tucson AZ 85713
Manufacturer (Section G)
SYNCARDIA SYSTEMS, LLC
1992 e silverlake road
tucson AZ 85713
Manufacturer Contact
aaron meier
1992 e silverlake road
tucson, AZ 85713
5205451234
MDR Report Key17006152
MDR Text Key315968038
Report Number3003761017-2023-00083
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 10/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number595000-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/17/2023
Initial Date FDA Received05/25/2023
Supplement Dates Manufacturer Received05/17/2023
Supplement Dates FDA Received10/14/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/22/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age32 YR
Patient SexMale
-
-