During the supraventricular tachycardia procedure, display and noise issues with the catheter resulted in procedural delays.The device was replaced and the procedure was completed with no adverse consequences to the patient.The ablation catheter was inserted into the atrium.After the catheter was connected, the three-dimensional shape of the catheter was abnormal, and the potential could not be displayed.The connection was checked, he tail wire was replaced, and the ampere, the amplifier and the dws were all restarted, but the issue could not be resolved.The catheter was replaced with a new one and the procedure ended smoothly with no adverse patient consequences.
|
One 4mm tip, large curl, safire ablation catheter was received for evaluation.The results of the investigation concluded that electrodes 1-4 met specifications of acceptable resistance values with no open or short circuits detected.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported noise and display issue remains unknown.
|