MEDTRONIC EUROPE SARL CLARIA MRI QUAD CRT-D SURESCAN; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
|
Back to Search Results |
|
Model Number DTMA1QQ |
Device Problems
Unable to Obtain Readings (1516); Inappropriate/Inadequate Shock/Stimulation (1574); Device Sensing Problem (2917)
|
Patient Problem
Shock from Patient Lead(s) (3162)
|
Event Date 04/30/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
Continuation of d10: pb1018, valve, implanted: (b)(6) 2022.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Event Description
|
It was reported that the patient became syncopal during an episode of fast supra ventricular tachycardia (svt).The cardiac resynchronization therapy defibrillator (crt-d) discriminators stopped withholding defibrillation therapy, the episode was classified as fascicular ventricular tachycardia (fvt), and the patient was shocked by the device.An interrogation of the device also noted invalid counter data and invalid atrial and ventricular histogram data.The patient was admitted to the hospital and the crt-d was reprogrammed to treat similar rhythms in the future as the physician felt the therapy was appropriate for the patient's condition.The crt-d remains in use.No further patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product event summary: the device was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory was performed and no anomalies were found.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|