Model Number 1456Q/75
Device Problem Use of Device Problem (1670)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/04/2023
Event Type  malfunction  
Manufacturer Narrative
Further information was requested but not received.
 
Event Description
It was reported the patient presented for initial procedure.During implant, the left ventricular lead was unable to be implanted.The physician elected not to use the lead.The patient condition was unknown.
 
Manufacturer Narrative
The reported event of unable to implant was not confirmed.A complete lead was returned in one piece for analysis.Electrical, visual, and s-curve evaluations were normal with no anomalies found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET LEAD SMALL-S, 75 CM
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17015169
MDR Text Key316664557
Report Number2017865-2023-21208
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734510141
UDI-Public05414734510141
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1456Q/75
Device Catalogue Number1456Q/75
Device Lot NumberA000123662
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/11/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/11/2023
Initial Date FDA Received05/26/2023
Supplement Dates Manufacturer Received06/02/2023
Supplement Dates FDA Received06/13/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/06/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age79 YR
Patient SexMale
-
-