Model Number 1458QL/86
Device Problems Over-Sensing (1438); Device Dislodged or Dislocated (2923)
Patient Problems Low Blood Pressure/ Hypotension (1914); Cardiac Perforation (2513); Pericardial Effusion (3271)
Event Date 05/09/2023
Event Type  Injury  
Event Description
It was reported that the patient presented with left ventricular (lv) lead dislodgement.During lv lead revision, a guidewire was found to have perforated into the pericardial space.Subsequently, the patient experienced hypotension and pericardial effusion.Pericardiocentesis was performed and the lv lead was explanted.Patient condition was stable, and the patient would continue to be monitored.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17015358
MDR Text Key316089103
Report Number2017865-2023-21262
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734510219
UDI-Public05414734510219
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1458QL/86
Device Catalogue Number1458QL/86
Device Lot NumberA000123964
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/09/2023
Initial Date FDA Received05/26/2023
Supplement Dates Manufacturer Received07/31/2023
Supplement Dates FDA Received07/31/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/13/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MRI CONDTENDRIL; OPTISURE MRI; UNIFY ASSURA
Patient Outcome(s) Required Intervention;
Patient Age83 YR
Patient SexMale
-
-