BOSTON SCIENTIFIC CORPORATION ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM; STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING
|
Back to Search Results |
|
Model Number 24657 |
Device Problems
Activation, Positioning or Separation Problem (2906); Material Deformation (2976)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/21/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that stent partial deployment occurred.Two 6x150, 130cm eluvia drug-eluting vascular stent systems were selected for use.During the procedure, both stents would not completely deploy.The handles, of both devices, were broken apart to pin and pull the stents for full deployment.The stents fully deployed at this point.There were no patient complications.And the patient is doing well post procedure.
|
|
Manufacturer Narrative
|
Device evaluated by mfr: returned product consisted of an eluvia self-expanding stent system.The outer sheath, tip, inner sheath and the remainder of the device were checked for damage.Visual examination revealed, that the handle is open.The inner liner is kinked 4.8cm from the tip.The middle sheath is kinked at the retainer and 10cm from the retainer.Microscopic examination revealed, no additional damages.The stent was deployed.And did not return for analysis.Inspection of the remainder of the device, revealed no other damage or irregularities.
|
|
Event Description
|
It was reported that stent partial deployment occurred.Two 6x150, 130cm eluvia drug-eluting vascular stent systems were selected for use.During the procedure, both stents would not completely deploy.The handles, of both devices, were broken apart to pin and pull the stents for full deployment.The stents fully deployed at this point.There were no patient complications and the patient is doing well post procedure.It was further reported that access was gained via ipsilateral approach.The target lesion was a difficult sub intimal with a high degree of calcification.It took hours to cross and prepare the lesion.A 0.035 inch guidewire was used.The stents exhibited stacking and an audible sound was heard in the handle prior to the rack pull.During the end of the deployment with the second stent, the stent elongated.
|
|
Manufacturer Narrative
|
B5: describe event or problem: additional information.Device evaluated by mfr: returned product consisted of an eluvia self-expanding stent system.The outer sheath, tip, inner sheath and the remainder of the device were checked for damage.Visual examination revealed that the handle is open.The inner liner is kinked 4.8cm from the tip.The middle sheath is kinked at the retainer and 10cm from the retainer.Microscopic examination revealed no additional damages.The stent was deployed and did not return for analysis.Inspection of the remainder of the device, revealed no other damage or irregularities.
|
|
Event Description
|
It was reported that stent partial deployment occurred.Two 6x150, 130cm eluvia drug-eluting vascular stent systems were selected for use.During the procedure, both stents would not completely deploy.The handles, of both devices, were broken apart to pin and pull the stents for full deployment.The stents fully deployed at this point.There were no patient complications and the patient is doing well post procedure.It was further reported that access was gained via ipsilateral approach.The target lesion was a difficult sub intimal with a high degree of calcification.It took hours to cross and prepare the lesion.A 0.035 inch guidewire was used.The stents exhibited stacking and an audible sound was heard in the handle prior to the rack pull.During the end of the deployment with the second stent, the stent elongated.
|
|
Manufacturer Narrative
|
Additional information added to device evaluated by mfr: further review determined that the clip was damaged.The device was reviewed, and it was concluded that the damage to the clip likely contributed to the deployment issue.As the thumbwheel was rotated, tension built up along the proximal inner and it caused the clip to start to deform.This suggests that procedural factors likely contributed to the creation of the damage that led to the deployment issue and stent deformation.Device evaluated by mfr: returned product consisted of an eluvia self-expanding stent system.The outer sheath, tip, inner sheath and the remainder of the device were checked for damage.Visual examination revealed that the handle is open.The inner liner is kinked 4.8cm from the tip.The middle sheath is kinked at the retainer and 10cm from the retainer.Microscopic examination revealed no additional damages.The stent was deployed and did not return for analysis.Inspection of the remainder of the device, revealed no other damage or irregularities.
|
|
Search Alerts/Recalls
|
|
|