BOSTON SCIENTIFIC CORPORATION ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM; STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING Back to Search Results
Model Number 24657
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Dyspnea (1816); Pain (1994); Dizziness (2194); Thrombosis/Thrombus (4440); Unspecified Heart Problem (4454); Restenosis (4576); Swelling/ Edema (4577)
Event Date 04/25/2023
Event Type  Injury  
Manufacturer Narrative
Patient identifier: subject (b)(6).Age at time of event: the patient was 68 years old at the time of study enrollment.
 
Event Description
S2444 elegance clinical study it was reported that on (b)(6) 2023, 570 days post-index procedure, the subject experienced symptoms related to stenosis in the left superficial femoral artery (sfa).Hospitalization and surgical intervention occurred in response to the event.The subject underwent treatment with three eluvia drug-eluting stents on (b)(6) 2021 as a part of the elegance clinical trial.The target lesion #001 was in the left proximal sfa, mid sfa, extending to distal sfa.The target lesion had a proximal reference vessel diameter of 7 mm, distal reference vessel diameter of 7 mm, and a lesion length of 260 mm.The target lesion was 100% stenosed and was thus classified as a tasc ii d lesion.Prior to target lesion treatment, pre-dilation was performed using a 4 mm x 200 mm non-bsc percutaneous transluminal angioplasty (pta) balloon.Treatment of target lesion was performed by the placement of study devices; two 6 mm x 120 mm eluvia drug-eluting stents, and one 7 mm x 80 mm eluvia drug-eluting stent.Following stent placement, post-dilation was performed using 5 mm x 150 mm, and 6 mm x 80 mm sterling otw pta balloons.The final residual stenosis was noted to be 0%.The same day, the subject was discharged on dual antiplatelet therapy.On (b)(6) 2023, subject visited the clinic with complains of shortness of breath, occasional dizziness, palpitations, edema, and a burning/stinging pain in her legs.On (b)(6) 2023, 578 days post-index procedure, the subject was admitted and underwent a peripheral arteriogram in the setting of critical limb ischemia of the left lower extremity with severely reduced abi at 0.3.Due to recurrent isr (in stent restenosis), the subject needed to be evaluated by vascular surgery for possible bypass surgery.In response to the event, surgery of left sfa was performed.Post procedure, thrombus was noted, and final residual stenosis was noted to be 0%.On (b)(6) 2023, subject had a vascular consultation where the subject complained of pain in the left calf and foot, which worsened at night.The subject reported walking occasionally relieved the pain, but could also increase claudication.The subject's left foot was noted to be discolored, specifically on the plantar forefoot region.On (b)(6) 2023, the event was considered resolved and the subject was discharged from the hospital on the same day.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM
Type of Device
STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jeff wallner
4100 hamline ave n
arden hills, MN 55112
6515811560
MDR Report Key17022342
MDR Text Key316193297
Report Number2124215-2023-25682
Device Sequence Number1
Product Code NIU
UDI-Device Identifier08714729876618
UDI-Public08714729876618
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P180011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Study,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 05/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date05/15/2023
Device Model Number24657
Device Catalogue Number24657
Device Lot Number0027518000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/02/2023
Initial Date FDA Received05/30/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/15/2021
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ELUVIA US, 6X120, 130 CM
Patient Outcome(s) Required Intervention; Hospitalization;
Patient SexFemale
Patient RaceWhite
-
-