MAUDE Adverse Event Report: PERFUSION SYSTEMS I.M.A. CANNULA; APPARATUS, AUTOTRANSFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PERFUSION SYSTEMS I.M.A. CANNULA; APPARATUS, AUTOTRANSFUSION Back to Search Results
Model Number 31001
Device Problems Obstruction of Flow (2423); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/07/2021
Event Type  malfunction  
Manufacturer Narrative
Device evaluation summary: visual inspection shows little to no signs of damage to the tip of the device.During the cleaning process there was no flow observed through the device.Reason for return was confirmed for no flow.Conclusion: after investigation at medtronic, complaint is confirmed for the blockage within the cannula tip.It is unknown what may have caused this occurrence.Review of the device history record found no abnormalities during manufacturing that would cause or contribute to the reported event.There were no adverse patient effects as a result of this incidence.Medtronic has made its supplier aware and will continue to monitor for future occurrences and trends.This regulatory report is being submitted as part of a retrospective review and remediation per d00953163 as part of a capa action.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information that during use, the customer reported the tip of the dlp i.M.A cannula was crushed and would not allow fluid flow.The device was replaced.There was no adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I.M.A. CANNULA
Type of Device
APPARATUS, AUTOTRANSFUSION
Manufacturer (Section D)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer (Section G)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key17029279
MDR Text Key316216307
Report Number2184009-2023-00358
Device Sequence Number1
Product Code CAC
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K810820
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 05/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/07/2023
Device Model Number31001
Device Catalogue Number31001
Device Lot Number2020100251
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/21/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/17/2021
Initial Date FDA Received05/31/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/07/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial