Model Number RA002-4545SLR70
Device Problems Corroded (1131); Mechanical Jam (2983)
Patient Problem Failure of Implant (1924)
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned for evaluation.The evaluation was performed by the london implant retrieval center (lirc).The root cause is unable to be determined at this time.A review of the device history records (dhr) confirmed the device met all quality inspections and specifications prior to release.If any additional information is provided, a supplemental report will be submitted.Report 2 of 2.
 
Event Description
An investigation report from an implant retrieval center was received and reported that during examination, the rod did not exert any force during force testing.Additionally, it was noted during surface damage grading that there was galling and corrosion.It was reported that the rod was removed as part of a planned removal procedure after the rod had fully extended; there was no adverse patient or user impact associated with this report.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC GENERATION 1.5 SPINAL BRACING AND DISTRACTION SYSTEM
Type of Device
GROWING ROD SYSTEM - MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr. ste 100
aliso viejo CA 92656
Manufacturer (Section G)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr. ste 100
aliso viejo CA 92656
Manufacturer Contact
geoff gannon
101 enterprise dr. ste 100
aliso viejo, CA 92656
8583448112
MDR Report Key17031705
MDR Text Key316245502
Report Number3006179046-2023-00234
Device Sequence Number1
Product Code PGN
UDI-Device Identifier00856719002145
UDI-Public856719002145
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K171791
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 05/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRA002-4545SLR70
Device Lot NumberA160222-02
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/05/2023
Initial Date FDA Received05/31/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured02/22/2016
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age2 YR
Patient SexFemale
-
-