It was reported that deflation issue occurred.During procedure, a 2.50 x 24mm synergy xd drug-eluting stent was advanced to treat the lesion.However, it was noticed that the balloon would not deflate with the guidewire.The procedure was completed with a different device.There were no patient complications reported and the patient status was stable.
|
Device evaluated by mfr.: synergy xd mr ous 2.50 x 24mm stent delivery system, catheter was returned for analysis.Visual, tactile, microscopic, dimensional, and functional analysis was performed on the device.Stent damaged was noted at the mid-section of the stent where stent struts were lifted and pulled off balloon.No signs of movement, stent was set between the proximal and distal markerbands.The crimped stent outer diameter was measured, and the result is within max crimped stent profile measurement.Balloon cones were reviewed, and no issues were noted.Balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.No issues identified with the hypotube shaft.No issues identified with the outer / mid-shaft sections or the inner lumen of the device.Bumper tip showed no signs of distal tip damage.The device could also be inflated to rbp and deflated with no issues.
|
It was reported that deflation issue occurred.During procedure, a 2.50 x 24mm synergy xd drug-eluting stent was advanced to treat the lesion.However, it was noticed that the balloon would not deflate with the guidewire.The procedure was completed with a different device.There were no patient complications reported and the patient status was stable.
|