MAUDE Adverse Event Report: COCHLEAR LTD CP1000 MAGNET; NUCLEUS 24 COCHLEAR IMPLANT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COCHLEAR LTD CP1000 MAGNET; NUCLEUS 24 COCHLEAR IMPLANT SYSTEM Back to Search Results
Model Number CP1000 MAGNET
Device Problem Inadequacy of Device Shape and/or Size (1583)
Patient Problems Erythema (1840); Necrosis (1971); Pain (1994); Skin Erosion (2075)
Event Type  Injury  
Manufacturer Narrative
This report is submitted on jun 1, 2023.
 
Event Description
Per the clinic, the patient experienced pain and redness at magnet site caused by the external magnet being too strong.This resulted in the development of a sore (initially thought to be a necrosis) at the implant site which was treated with a topical antibiotic and stopping use of the device until the area healed.The magnet was changed from a 2 to a ½ strength.The implant remains in-situ.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CP1000 MAGNET
Type of Device
NUCLEUS 24 COCHLEAR IMPLANT SYSTEM
Manufacturer (Section D)
COCHLEAR LTD
1 university avenue
macquarie university, nsw 2109
AS  2109
Manufacturer (Section G)
COCHLEAR LTD
1 university avenue
macquarie university, nsw 2109
AS   2109
Manufacturer Contact
roberta loveday
1 university avenue
macquarie university, nsw 2109
AS   2109
MDR Report Key17035571
MDR Text Key316278089
Report Number6000034-2023-01722
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970051
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/31/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberCP1000 MAGNET
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/13/2023
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexFemale
-
-