• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH UNK - NAILS: TFNA; NAIL,FIXATION,BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH UNK - NAILS: TFNA; NAIL,FIXATION,BONE Back to Search Results
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Pain (1994)
Event Date 05/16/2023
Event Type  Injury  
Manufacturer Narrative
Product complaint #: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.This report is for an unk - nails: tfna/unknown lot.Part and lot number are unknown.Without the specific part number; the udi number and 510-k number is unknown.D9: complainant part is not expected to be returned for manufacturer review/investigation.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.This complaint involves three (3) devices.H3/h6: the photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device from attachment (img_2389.Jpeg).Visual analysis of the photo revealed that unk - nails: tfna had no signs of issue or defect.X-ray evidence provided shows a nail and helical blade screw construct at the left femur.Device does not appear to have any defect or malfunction.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was not confirmed for unk - nails: tfna.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot a manufacturing record evaluation cannot be performed due to lot number being unknown.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that on (b)(6) 2023, the surgeon removed an unknown tfna construct from an unknown patient.This hardware was removed because of pain.The nail construct was successfully removed on (b)(6) 2023 and the surgeon used norian to backfill where the implant was removed.This complaint involves three (3) devices.This report is for one (1) unk - nails: tfna.This is report 2 of 3 for complaint (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNK - NAILS: TFNA
Type of Device
NAIL,FIXATION,BONE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key17042699
MDR Text Key316384044
Report Number8030965-2023-06907
Device Sequence Number1
Product Code JDS
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/16/2023
Initial Date FDA Received06/01/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
UNK - NAIL HEAD ELEMENTS: TFNA HELICAL BLADE.; UNK - SCREWS: NAIL DISTAL LOCKING.
Patient Outcome(s) Required Intervention;
-
-