• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION MOMENTUM CRT-D; IMPLANTABLE DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION MOMENTUM CRT-D; IMPLANTABLE DEFIBRILLATOR Back to Search Results
Model Number G124
Device Problems Failure to Convert Rhythm (1540); Inaccurate Synchronization (1609)
Patient Problem Arrhythmia (1721)
Event Date 05/12/2023
Event Type  Injury  
Event Description
It was reported that the patient with this cardiac resynchronization therapy defibrillator (crt-d) entered in an intrinsic ventricular tachycardia episode.The device delivered appropriately anti tachy therapy (atp), however, this accelerated the rhythm.The device delivered seven shocks in order to end the episode, however, they were unsuccessful and the therapy got exhausted.The patient was administered medication to control the arrhythmia and was admitted for hospitalization.This device remains in service.No further adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOMENTUM CRT-D
Type of Device
IMPLANTABLE DEFIBRILLATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key17043352
MDR Text Key316351510
Report Number2124215-2023-27920
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526589102
UDI-Public00802526589102
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S436
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 06/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/24/2023
Device Model NumberG124
Device Catalogue NumberG124
Device Lot Number503535
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/12/2023
Initial Date FDA Received06/01/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/18/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other; Hospitalization;
Patient Age72 YR
Patient SexMale
-
-