Model Number 470057-08
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/14/2022
Event Type  malfunction  
Manufacturer Narrative
Intuitive surgical, inc.(isi) received the instrument involved with this complaint and completed the device evaluation.Failure analysis confirmed the customer reported complaint.The attached endoscope adapter (aea) was damaged or had a friction issue.The endoscope cable integrity exhibited visual damage.
 
Event Description
It was reported that during a da vinci-assisted surgical procedure, the endoscope moved in the opposite direction to the joystick control.The procedure was completed with no reported injury.Intuitive surgical, inc.(isi) followed up and obtained the following additional information: the wire insulation of the endoscope was damaged/defective.The surgical team confirmed that there was no issue with the joystick of the endoscope.The procedure was successfully completed with no reported injury and no delay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NONE
Type of Device
ENDOSCOPE PLUS
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 
4085232100
MDR Report Key17046120
MDR Text Key316481984
Report Number2955842-2023-15538
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier00886874116555
UDI-Public(01)00886874116555
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K191736
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign
Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial
Report Date 02/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/02/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number470057-08
Device Catalogue Number470057
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/22/2022
Date Manufacturer Received02/22/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/03/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
DA VINCI INSTRUMENTS AND ACCESSORIES
Patient SexMale
-
-