Model Number 173016
Device Problems Component or Accessory Incompatibility (2897); Detachment of Device or Device Component (2907); Human-Device Interface Problem (2949)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/11/2023
Event Type  malfunction  
Event Description
According to the reporter, during a total laparoscopic hysterectomy procedure, when closing of vaginal cuff, the suturing device was requested to close and the surgeon used a suture.The surgical technologist was attempting to load the sutures into the device, but it was not working properly.It was extremely hard to pull the bars down to capture the needle in the suturing device.The needle would fall out of device after toggling even though the arms were down.The surgical technologist requested an additional suturing device.The second device experienced technical issues when attempting to transfer the needle from side to side; however the surgical technologist was able to force it from left and right to avoid opening a third device.There were multiple sutures wasted because of the malfunction of the device, which caused a delay for 10-15 minutes during the procedure.There was no patient injury.
 
Manufacturer Narrative
Concomitant medical products: 173016 endo stitch instrument (lot#: j2e2944ey); unknown endo stitch sulu (lot#: unknown); unknown endo stitch sulu (lot#: unknown) evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the needle fell out of the jaws of the device and the device was difficult to load.The device was difficult to toggle.The reported issues could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO STITCH
Type of Device
ENDOSCOPIC TISSUE APPROXIMATION DEVICE
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR  0101
Manufacturer (Section G)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR   0101
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key17051268
MDR Text Key316445087
Report Number9612501-2023-00907
Device Sequence Number1
Product Code OCW
UDI-Device Identifier10884521100008
UDI-Public10884521100008
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K934738
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 06/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number173016
Device Catalogue Number173016
Device Lot NumberJ2F0156EY
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/29/2023
Initial Date Manufacturer Received 05/15/2023
Initial Date FDA Received06/02/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/02/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexFemale
-
-