Correction: the initial report was incorrectly submitted with a cfn-based manufacturer report (mfr) number: (b)(4).The mfr number should have been fei-based with the 10-digit fei being (b)(4).Manufacturer's investigation conclusion: the reported event of a flow issue was confirmed via the log file.The centrimag 2nd generation primary console (serial #: (b)(6)) was returned for analysis and a log file was downloaded for review and showed relevant events spanning approximately 4 days (23 apr 2023 ¿ 25 apr 2023, 18 may 2023 per time stamp).The console was operating a motor at a speed of ~3600 rpm with a flow of ~3.5 lpm.Beginning on 24 apr 2023 at 03:09, the flow dropped below 1 lpm and continued to decrease to be a negative value.A ¿flow below minimum: f3¿ alarm activated.The console was shutdown at 03:26.The console was power cycled several times before returning to abbott.There were no other notable events active in the log file.Pump operation was not affected.The console was connected to the returned and associated centrimag motor and flow probe and was run for several days on a mock loop as well as tested independently of the flow probe and motor.The reported event was unable to be reproduced.The console was functionally tested and passed all tests.The root cause for the reported event was unable to be conclusively determined through this analysis.The 2nd generation centrimag system operating manual, rev.L, section 4 entitled "warnings & precautions" warns "one additional 2nd generation centrimag primary console, motor and flow probe are required as backup system in the immediate vicinity of each patient whenever the centrimag or pedivas blood pump is used.The backup console must be connected to the backup motor and to the backup flow probe, have a battery charge sufficient for at least one hour of operation, be connected to ac power (except during transport) and be immediately available should the main console, motor or flow probe experience a malfunction." the 2nd generation centrimag system operating manual, rev.L, section 10 entitled "emergency and troubleshooting" states that "the recommended practice whenever there is a 2nd generation centrimag primary console or motor malfunction is to replace the console and motor as a set.Remove the blood pump from the malfunctioning motor and console and place the blood pump in the backup motor and console to continue patient support.Do not exchange individual motors or individual consoles during patient support.The device history records were reviewed for the centrimag 2nd generation primary console (serial #: (b)(6)) and the console was found to pass all manufacturing and quality assurance (qa) specifications.No further information was provided.The manufacturer is closing the file on this event.
|