• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT GALLANT HF; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT GALLANT HF; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CDHFA500Q
Device Problems Connection Problem (2900); Contamination of Device Ingredient or Reagent (2901)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/17/2023
Event Type  malfunction  
Event Description
During the initial implant, a small spring was observed in the device header and the lead was unable to be connected to the device.The device was explanted and replaced to resolve the event.The patient was stable.
 
Manufacturer Narrative
The reported event of dislodged contact spring in the header was confirmed.A dislodged spring in the atrial chamber of the header was observed.The investigation found an issue related to manufacturing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT HF
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17053756
MDR Text Key316505177
Report Number2017865-2023-22274
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067032010
UDI-Public05415067032010
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCDHFA500Q
Device Catalogue NumberCDHFA500Q
Device Lot NumberP000161996
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/17/2023
Initial Date FDA Received06/02/2023
Supplement Dates Manufacturer Received08/31/2023
Supplement Dates FDA Received09/06/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/28/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-