MAUDE Adverse Event Report: BIOSENSE WEBSTER INC THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER; CARDIAC ABLATION PERCUTANEOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER; CARDIAC ABLATION PERCUTANEOUS CATHETER Back to Search Results
Model Number D134805
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/09/2023
Event Type  malfunction  
Event Description
It was reported that a patient underwent an atrial tachycardia (at) right ablation procedure with two thermocool® smart touch® sf bi-directional navigation catheters and an issue with bad/no ecg on all channels occurred.It was reported that there was noise ablation distal and proximal and all 12-leads on the mapping, recording, and anesthesia monitor.They switched the ecg trunk cable and swapped the leads cables as well, but the issue did not resolve.The cable was replaced without resolution.The catheter was replaced, however, the issue continued.Catheter and cable.Grounding pad on the patient.The generator was replaced, the pump was replaced, and the issue did not resolve.The generator to pump cable was replaced but the noise issue was still there.The spu was not in use, so they disconnected the spu power cord, but the issue was still there.Restarted the piu and magnet but that made no difference.Brought in a new piu magnet and other part of the system from another room and the issue was still there.¿when in with a low wattage signal¿ and the anesthesia team noticed that the noise in and out matches the patient's respirations.While looking at fluoro and noticed when the patient defibrillator leads were closed to the catheter's ablation signals and the leaked signal was entering the patient's body and causing the noise issue.The noise was on all body surface (bs) and ablation intracardiac (ic) on both carto and recording system.Cs ic (also through our piu) had no noise on either carto or recording system.During radiofrequency (rf), the physician had no intact bs ecg to monitor patient¿s heart rate (noise washed out all bs ecgs).With cessation of rf, all bs ecgs returned to normal.The case continued and ended successfully.No adverse patient consequence was reported.The bad/no ecg on all channels is mdr reportable.
 
Manufacturer Narrative
If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Biosense webster manufacturer's reference number (b)(4), has two reports: (1) mfr # (b)(4) for product code d134805 (thermocool® smart touch® sf bi-directional navigation catheter).(2) mfr # 2029046-2023-01212 for product code d134805 (thermocool® smart touch® sf bi-directional navigation catheter).
 
Manufacturer Narrative
The biosense webster inc.(bwi) product analysis lab received the device for evaluation on 12-jun-2023.The device evaluation was completed on 05-jul-2023.It was reported that a patient underwent an atrial tachycardia (at) right ablation procedure with two thermocool® smart touch® sf bi-directional navigation catheters and an issue with bad/no ecg on all channels occurred.Device evaluation details: the device was returned to biosense webster (bwi) for evaluation.Visual inspection and electrical test of the returned device were performed following bwi procedures.Visual analysis revealed no damage or anomalies on the device.An electrical test was performed, and no electrical issues were found.During product analysis, the lot number was verified to be 31025440l, based on the electronically erasable programmable read only memory (eeprom) file.Therefore, the d4.Lot, d4.Expiration date and h4.Device manufacture date have been updated.A manufacturing record evaluation was performed for the finished device number lot 31025440l and no internal actions related to the complaint were found during the review.The electrical issue reported by the customer could not be replicated during the product investigation; other issues or circumstances may have occurred during the usage of the device that compromised its performance.The instructions for use contain the following recommendation: ecg noise is typically generated as the result of the improper connection of the body surface ecg patch to the patient.This noise is the most significant during ablation.To resolve this situation, verify the proper connection.It is recommended to turn off the notch filter for this verification.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Biosense webster manufacturer's reference number pc-001351334 has two reports: (1) mfr # 2029046-2023-01210 for product code d134805 (thermocool® smart touch® sf bi-directional navigation catheter).(2) mfr # 2029046-2023-01212 for product code d134805 (thermocool® smart touch® sf bi-directional navigation catheter).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER
Type of Device
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key17056464
MDR Text Key316864031
Report Number2029046-2023-01210
Device Sequence Number1
Product Code LPB
UDI-Device Identifier10846835010183
UDI-Public10846835010183
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberD134805
Device Catalogue NumberD134805
Device Lot Number31025440L
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/09/2023
Initial Date FDA Received06/02/2023
Supplement Dates Manufacturer Received06/12/2023
Supplement Dates FDA Received07/11/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/31/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
THMCL SMTCH SF BID, TC, D-F.; UNK BRAND CABLE.; UNK BRAND CABLE.; UNK BRAND CABLE.; UNK GENERATOR.; UNK PUMP.
-
-