• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOSENSE WEBSTER (ISRAEL) LTD. CARTO 3; COMPUTER, DIAGNOSTIC, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER (ISRAEL) LTD. CARTO 3; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Model Number FG540000
Device Problem Defective Component (2292)
Patient Problem Insufficient Information (4580)
Event Date 04/13/2023
Event Type  malfunction  
Event Description
During an atrial fibrillation ablation procedure and mapping of the left atrium, there was a failure of ecg 1 in the piu of the carto mapping system causing a delay and switch to a portable system.Manufacturer response for mapping system, carto 3 system (per site reporter).Device was repaired.(b)(4).Findings: defective part - backplane card.Defective cable - bodysurface electrograms.Defective part - ecg card.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARTO 3
Type of Device
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
BIOSENSE WEBSTER (ISRAEL) LTD.
31 technology drive, suite 200
irvine CA 92618
MDR Report Key17057292
MDR Text Key316491765
Report Number17057292
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberFG540000
Device Catalogue NumberFG-5400-00
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA05/23/2023
Event Location Hospital
Date Report to Manufacturer06/05/2023
Initial Date Manufacturer Received Not provided
Initial Date FDA Received06/05/2023
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age28835 DA
Patient SexFemale
-
-