Model Number 24653
Device Problems Premature Activation (1484); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/29/2023
Event Type  malfunction  
Event Description
It was reported that the stent prematurely deployed.This 7x40, 130 cm eluvia drug-eluting vascular stent was selected for use in an endovascular therapy procedure.The 100% stenosed lesion was located in the mildly tortuous and severely calcified superficial femoral artery.During delivery, the tip of the stent protruded slightly from the shaft due to calcification on the lesion.Use of the device was discontinued, and it was removed from the sheath.It was noted the safety lock had been on.Another pre-dilation was performed, and the procedure was completed with a different sized eluvia stent.There was no patient injury.
 
Manufacturer Narrative
Device eval by manufacturer: upon receipt at our post market quality assurance laboratory, this 7x40, 130 cm eluvia drug-eluting vascular stent system was inspected for damage.Visual examination revealed a kink to the outer sheath at the nosecone.The stent was partially deployed 8mm from the distal end of the middle sheath.Microscopic examination revealed no additional damages.The yellow thumbwheel lock and rack were still in the manufactured position.There was blood inside the middle sheath.Inspection of the remainder of the device, revealed no other damage or irregularities.Product analysis confirmed the stent was partially deployed.
 
Event Description
It was reported that the stent prematurely deployed.This 7x40, 130 cm eluvia drug-eluting vascular stent was selected for use in an endovascular therapy procedure.The 100% stenosed lesion was located in the mildly tortuous and severely calcified superficial femoral artery.During delivery, the tip of the stent protruded slightly from the shaft due to calcification on the lesion.Use of the device was discontinued, and it was removed from the sheath.It was noted the safety lock had been on.Another pre-dilation was performed, and the procedure was completed with a different sized eluvia stent.There was no patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM
Type of Device
STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key17060305
MDR Text Key316898724
Report Number2124215-2023-27500
Device Sequence Number1
Product Code NIU
Combination Product (y/n)Y
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/11/2024
Device Model Number24653
Device Catalogue Number24653
Device Lot Number0028764443
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/30/2023
Initial Date FDA Received06/05/2023
Supplement Dates Manufacturer Received07/12/2023
Supplement Dates FDA Received07/24/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/11/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-