• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH TROCHANTERIC NAIL KIT, TI GAMMA3® Ø11X180MM X 125°; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH TROCHANTERIC NAIL KIT, TI GAMMA3® Ø11X180MM X 125°; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 3125-1180S
Device Problem Positioning Failure (1158)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/06/2023
Event Type  Injury  
Manufacturer Narrative
Based on the available information the device will not be returned therefore an evaluation of the device cannot be performed.Should additional information become available, it will be provided in a supplemental report.
 
Event Description
As reported: "after placement of the nail and cephalic screw, impossible placement of the locking screw with an audible noise as soon as the screw was put in contact without starting to screw.The team was forced to remove all the material and re-install another nail.When the nail was tested outside the patient, the locking screw was still not inserted.Additionally, there was a delay of 20 to 30 minutes.".
 
Manufacturer Narrative
Please note correction to d3 (manufacturer entity) and d4 (lot #).The reported event could not be confirmed, since the device was not returned for evaluation and no other evidences were provided.The device inspection was not possible as the product was not returned for investigation.The device history record could not be reviewed because the affected device was not returned, and the lot number was not communicated.A review of the labeling did not indicate any abnormalities.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If device is returned or any further information is provided, the investigation report will be reassessed.
 
Event Description
As reported: "after placement of the nail and cephalic screw, impossible placement of the locking screw with an audible noise as soon as the screw was put in contact without starting to screw.The team was forced to remove all the material and re-install another nail.When the nail was tested outside the patient, the locking screw was still not inserted.Additionally, there was a delay of 20 to 30 minutes.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TROCHANTERIC NAIL KIT, TI GAMMA3® Ø11X180MM X 125°
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key17062148
MDR Text Key316573640
Report Number0009610622-2023-00191
Device Sequence Number1
Product Code HSB
UDI-Device Identifier04546540375070
UDI-Public04546540375070
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K034002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 07/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3125-1180S
Device Catalogue Number31251180S
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/09/2023
Initial Date FDA Received06/05/2023
Supplement Dates Manufacturer Received06/22/2023
Supplement Dates FDA Received07/14/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-