• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOVASCULAR LLC VASOVIEW HEMOPRO 2; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOVASCULAR LLC VASOVIEW HEMOPRO 2; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number VH-4000
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/11/2023
Event Type  malfunction  
Event Description
The hospital reported that during an endoscopic vein harvesting procedure, when activating the jaws of vasoview hemopro 2, the jaws burned through the insulation.A new kit was opened to finish the procedure.No patient effects were reported.
 
Manufacturer Narrative
Trackwise id (b)(4).The device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.
 
Manufacturer Narrative
Trackwise#: (b)(4).The device was returned to the factory for evaluation on 05/24/2023.An investigation was conducted on 05/31/2023.A visual inspection was conducted.Sings of clinical use and evidence of charred material was observed on the heater wire.The heater wire was flexed away from the hot jaw at the center due to normal clinical use.There were no visual defects observed on the clear silicone insulation on both the cold and hot jaws.No melting or thermal defects were observed.Based on the returned condition of the device as well as the evaluation results, the reported failure "thermal decomposition of device" was not confirmed.The lot # 3000279125 history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VASOVIEW HEMOPRO 2
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 
MDR Report Key17078154
MDR Text Key316678423
Report Number2242352-2023-00452
Device Sequence Number1
Product Code GEI
UDI-Device Identifier00607567700406
UDI-Public00607567700406
Combination Product (y/n)N
PMA/PMN Number
K101274
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 06/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/07/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberVH-4000
Device Catalogue NumberVH-4000
Device Lot Number3000279125
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/24/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/16/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/09/2022
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-