• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CERNER CORPORATION CERNER MILLENNIUM BLOOD BANK TRANSFUSION; BLOOD ESTABLISHMENT COMPUTER SOFTWARE AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CERNER CORPORATION CERNER MILLENNIUM BLOOD BANK TRANSFUSION; BLOOD ESTABLISHMENT COMPUTER SOFTWARE AND ACCESSORIES Back to Search Results
Device Problems Application Program Problem (2880); Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/16/2023
Event Type  Other  
Event Description
On (b)(6) 2023, a problem was reported to (b)(6) that a previous complaint regarding how entering donor directed blood products into inventory was giving an option to choose a donation type of directed donor and autologous was a risk and it should always default to directed donor.While performing the current upgrade validations we noticed that the problem was not corrected.This is an excerpt from communication between our it(information technology) and (b)(6), it was such a pleasure meeting with you all this afternoon.I have taken some time to document the different scenarios we tested.Please review the following and make any corrections necessary.I am also attaching a spreadsheet with test cases and observations.Lots of copy paste so it may need some edits on my part.Please review the following summary from our meeting and testing: when directed product barcode is scanned with both allogeneic and directed products built in system, pop-up message is being displayed for user to select the product rather than the directed product being defaulted.In another scenario, with the exact same product type scanned as above, there is no pop displayed if the aborh type embedded with directed use only barcode is scanned.However, the product that defaults is not directed product but the allogeneic product with donation type of directed crossover.Wrong product name with no indication of the mismatch donation type embedded for aborh field (directed use only) and product type field (directed crossover).When only allogeneic product is built and the directed product is scanned, the allogeneic product is defaulted with donation type of "directed crossover" rather than the pop-up "product is not found; product is not aliased" as is expected.Please let me know if this accurately represents the issue and i will forward our investigation to the correct team.Thank you so much for your patience! summary: regulatory compliance issue (b)(4), sr1-(b)(4) description: we downloaded (b)(4) to fix a compliance issue that came in with code level 2018.11 code.We submitted an obligation with the regulatory requirements.This package did not address that issue.I am resubmitting the documentation to be taken back for an obligation for a fix.I am attaching screenshots where previous code worked as designed and was in alignment with regulatory requirements and screenshots after the package in m1805.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CERNER MILLENNIUM BLOOD BANK TRANSFUSION
Type of Device
BLOOD ESTABLISHMENT COMPUTER SOFTWARE AND ACCESSORIES
Manufacturer (Section D)
CERNER CORPORATION
MDR Report Key17079749
MDR Text Key316788015
Report NumberMW5118230
Device Sequence Number1
Product Code MMH
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received06/06/2023
Patient Sequence Number1
-
-