Model Number N/A |
Device Problems
Failure to Cut (2587); Calibration Problem (2890)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/04/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
This complaint is recorded by zimmer biomet under (b)(4).A follow up/ final report will be submitted once investigation is complete.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
It was reported that the unit needed calibration since the unit was not taking adequate skin graft.There was no harm or delay reported.No adverse events were reported as a result of this malfunction.Diligence is in progress.
|
|
Manufacturer Narrative
|
Following sections were updated or corrected : b3, b5, d2, g1, g3, g6, h1, h2 and h6.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
Additional information received determined that no additional graft was needed.Due diligence is complete and no further information is available.
|
|
Manufacturer Narrative
|
Review of the most recent repair record determined the control bar was out of calibration and the thickness control lever's ball detent was worn.The control bar, thickness control lever, machined head, bearings, eccentric shaft, and adjustment cams were replaced to resolve the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
There is no additional event information available.
|
|
Manufacturer Narrative
|
If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
There is no additional event information available.
|
|