MAUDE Adverse Event Report: OAKDALE LEVEL 1 HOTLINE FLUID WARMER; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OAKDALE LEVEL 1 HOTLINE FLUID WARMER; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number CON-HL-90
Device Problem Defective Component (2292)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Event Description
It was reported, that two of the test buttons on the side of the device were not triggering the alarms.Additionally, the front of the case was cracked.No patient harm or involvement reported.
 
Manufacturer Narrative
B3: date of event; no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr, 803.56 when required.
 
Manufacturer Narrative
Device evaluation: one device was received for investigation.The visual inspection showed corroded drain fitting, a cracked tank cover, outdated pcb and power switch, faded line cord, and damaged lcd screen.During the functional testing it was found that the buttons on the membrane switch aren't working, which confirmed the customer complaint.The root cause was found to be a faulty or worn out membrane switch.No action was taken due to the condition of the device.It is deemed beyond economical repair and will be scrapped.A manufacturing dhr review was not performed because the device is beyond a year from its manufacture date of 2009 and there was no indication of a manufacturing defect during the investigation.Service history review identified this device has not been in for service previously.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17082380
MDR Text Key316751486
Report Number3012307300-2023-06085
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/07/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCON-HL-90
Device Catalogue NumberHL-90
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received06/09/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/06/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-