The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd module with drive pcb ccd module out of focus.Based on the result, we concluded that it was caused due to the moisture condensation in the ccd module with drive pcb.In addition, our technician confirmed that the lcb (light carrying bundle) broken, the remote control buttons leak, the remote control buttons cut, the light guide cable for control body faulty connection, and the light guide cable for prong faulty connection; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report and/or the risk analysis results, it was evaluated to submit mdr.
|