DEPUY MITEK LLC US COOLPULSE 90 ELECTRODE WITH HAND CONTROLS; ENDOSCOPIC ELECTROSURGICAL ELECTRODE, BIPOLAR, SINGLE USE
|
Back to Search Results |
|
Model Number 228147 |
Device Problem
Device Remains Activated (1525)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/20/2022 |
Event Type
malfunction
|
Event Description
|
It was reported by the sales rep in brazil that during an unknown procedure on (b)(6)2022, it was observed that the coolpulse 90 electrode with hand controls device presented failure during use.According to the report, it was in ablation every moment after pause.Another like device was used to complete the procedure.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
|
|
Manufacturer Narrative
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: udi: (b)(4).E3: reporter is a j&j sales representative.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|
|
Manufacturer Narrative
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: d9, h3, h6: the actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available fo the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary = > the product was returned to mitek for evaluation.Mitek then conducted visual inspection of the device received.The complaint device was received and evaluated in juarez lab.Visual inspection revealed that the device was broken, a deformation curve was found in the distal part of the shaft, a partial part of the shaft was still attached in the handle.The power cord as well as the suction tube were found cut, the cut pieces were not returned.The device tip shows rests of foreign matter, presumably biological matter.Also, the metal tip was found damaged.A manufacturing record evaluation was performed for the finished device u2110084, and no non-conformances were identified.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.Based on the fact that the functional test of the device could not be performed due to the condition in which it was received, this complaint cannot be confirmed and it is not possible to determine a possible root cause for the issue experienced by the customer.The possible root cause of the physical damage found in the device can be attributed to procedural variables, such as handling of the device or interaction of the product during the procedure; a force greater than the design specification may have been applied to the device during use, causing the tip to be damaged and the shaft to bend and consequently it breakage, however, this cannot be conclusively determined.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
|
|
Manufacturer Narrative
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatchwill be filed as appropriate.H10 additional narrative: investigation summary : the complaint device was not returned after multiple attempts for device return, therefore unavailable for a physical evaluation.Since the complaint device was not returned, we cannot determine a root cause for the reported failure.If additional information or the device is received in the future, we will reopen the complaint and perform the investigation as appropriate.A manufacturing record evaluation was performed for the finished device u2110084, and no non-conformances were identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
|
|
Search Alerts/Recalls
|
|
|