• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. DERMATOME AN HANDPIECE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. DERMATOME AN HANDPIECE Back to Search Results
Model Number N/A
Device Problems Loose or Intermittent Connection (1371); Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/03/2023
Event Type  malfunction  
Manufacturer Narrative
This event is recorded with zimmer biomet under (b)(4).This medwatch is being filed as an initial / final report based on information discovered during the device evaluation.Review of the most recent repair record determined the control bar was not in the correct position and the thickness control lever was loose.The head, neck, control bar, control lever, motor, and various other parts were scrapped and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that the dermatome needed to be calibrated for preventative maintenance.The device was sent in for repair.During test and calibration, the unit was found to be in need of repair.The rpms were in specification.The control bar was not in the correct position.The calibration was in at all readings.The thickness control lever was loose.Due diligence is complete.No additional information is available at this time.
 
Event Description
There is no additional event information available.
 
Manufacturer Narrative
If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DERMATOME AN HANDPIECE
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key17090982
MDR Text Key316970159
Report Number0001526350-2023-00557
Device Sequence Number1
Product Code GFD
UDI-Device Identifier00889024464834
UDI-Public(01)00889024464834(11)180801(10)64092586
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/08/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number88710100
Device Lot Number64092586
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/10/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/04/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-