Model Number 307F3 |
Device Problems
Adverse Event Without Identified Device or Use Problem (2993); Protective Measures Problem (3015)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/22/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Event Description
|
It was reported that during a cryo ablation procedure, a system notice was received indicating that the injection was stopped because the cryomapping temperature exceeded the preset value by more than fifteen degrees.The case was aborted and the patient was under general anesthesia.Test ablations were performed and the console appeared to be controlling pressure and flow in the expected ranges.A preventative maintenance visit was recommended. no patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
E: initial reporter - updated product event summary: data files were returned and analyzed.Patient files showed one application was performed with the 307f3 elec trophysiology (ep) catheter with lot 15685 on the date of the event.The patient files showed system notice 50016 "indicating that the injection has stopped because the cryomapping temperature exceeded the preset value by more than 15 degrees." in conclusion, the reported system notice 50016 was confirmed.The reported clinical issue (aborted under general anesthesia) cannot be assessed through data analysis.The physical product, 307f3 catheter with lot 15685 was discarded and was not returned for analysis.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|