A facility reported that at the start of the case, the surgeon was setting up a cerelink icp sensor but could not get the p-zero screen on the icp express monitor.The microsensor was attempted to be zeroed for approximately 30 minutes before integra¿s account manager was contacted.The account manager was supporting over the phone for approximately 15 minutes before the insertion procedure was abandoned and the spinal procedure commenced (the patient was scheduled for a spine surgery after sensor placement).At the end of the spinal procedure, the account manager was onsite to test the consoles and advise on successful insertion.Troubleshooting taken: a demo sensor was used by the integra account manager to test the two icp express monitors with their grey cables.One worked, one didn¿t.The account manager then swapped grey cables between monitors and found one cable was not working (cable 826636 lot 105512).Once the faulty cable was isolated and removed, medical staff were able to successfully complete zeroing steps.No issues after insertion.No variation in icp was noted, and icp sensor was removed after 24 hours.Reason for sensor placement; fall from height with spinal injuries and suspect bleeding on the brain.No issues were detected overnight and the icp was removed the following day.
|