• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY GMBH TUBING SET; OXYGENATOR, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY GMBH TUBING SET; OXYGENATOR, CARDIOPULMONARY BYPASS Back to Search Results
Model Number HLS SET ADVANCED
Device Problems Pressure Problem (3012); Temperature Problem (3022)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/08/2023
Event Type  malfunction  
Manufacturer Narrative
A follow-up medwatch will be submitted when additional information becomes available.
 
Event Description
It was reported that the customer stated that the device was giving alternating error messages in the status bar at the top of the cardiohelp screen.The error messages were arterial pressure (part) and arterial temperature (tart) sensor defective.The customer exchanged the hls cable, but the failure was not solved.The customer switched the application modus.The error messages were resolved.The failure occurred during treatment.No harm to any person has been reported.Complaint id# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TUBING SET
Type of Device
OXYGENATOR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
Manufacturer (Section G)
JULIA KAPFENBERGER
neue rottenburger strasse 37
hechingen
Manufacturer Contact
neue rottenburger strasse 37
hechingen 
MDR Report Key17105088
MDR Text Key317649642
Report Number8010762-2023-00277
Device Sequence Number1
Product Code DTZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112360
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHLS SET ADVANCED
Device Catalogue Number701069078 (SET)
Device Lot Number3000293725 (MODUL)
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/08/2023
Initial Date FDA Received06/12/2023
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Type of Device Usage Initial
Patient Sequence Number1
-
-