ZIMMER ORTHOPAEDIC MFG. LTD. ARTICULAR SURFACE FIXED BEARING (CPS) RIGHT 12 MM HEIGHT; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Device Problem Difficult to Insert (1316)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/19/2023
Event Type  malfunction  
Manufacturer Narrative
(b)(4) multiple mdr reports were filed for this event, please see associated reports: 3007963827-2023-00153 and 0001822565-2023-01560.D10 medical devices: tibia fixed cemented right size f catalog#: 42542007502 lot#: 65845201 unknown femoral catalog#: ni lot#: ni g2 foreign source: japan customer has indicated that the product will not be returned to zimmer biomet for investigation, as it remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that during a knee procedure, the tibial insert would not seat on the tibial tray.A impactor was used to seat the tibial insert into the tray.No adverse events have been reported as a result of the malfunction.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Reported event was unable to be confirmed due to limited information received from the customer.Device history record was reviewed and no discrepancies were found.Review of complaint history found no additional related issues for this item and the reported part and lot combination.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICULAR SURFACE FIXED BEARING (CPS) RIGHT 12 MM HEIGHT
Type of Device
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
Manufacturer (Section G)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI  
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17117385
MDR Text Key317206402
Report Number3007963827-2023-00153
Device Sequence Number1
Product Code MBH
UDI-Device Identifier00889024244580
UDI-Public(01)00889024244580(17)260317(10)64981271
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K123459
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number42522600712
Device Lot Number64981271
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/19/2023
Initial Date FDA Received06/13/2023
Supplement Dates Manufacturer Received10/16/2023
Supplement Dates FDA Received10/19/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/08/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-