LIVANOVA DEUTSCHLAND GMBH S5 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Catalog Number 10-80-00
Device Problem Failure to Auto Stop (2938)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/12/2023
Event Type  malfunction  
Event Description
Livanova deutschland received a report that, during priming a s5 roller pump, even if stopped by the user, it started again in the opposite direction for a while and then stopped.It was an intermittent failure; it happened three (3) times.There was no patient involvement.
 
Manufacturer Narrative
A.1.-a.5.There was no patient involvement.H10: livanova deutschland manufactures the s5 roller pump.The incident occurred in france.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Manufacturer Narrative
H10: the unit was sent to the manufacturing site for further investigation.The reported issue was confirmed and the motor control board (hms) was replaced to solve it.Subsequent functional verification testing was completed without further problems and the unit was returned to service.The hall sensors of the pump motor, which detects the position of the rotor's rotating magnetic field, are connected to the board involved via a socket.Consequently, based on technical intervention reported above, it cannot be ruled out that this hms board failed and did not allow for correct signal processing.The wearing of electro-mechanical device can be originated by the specific use condition at customer site and may have contributed to the event.The pump involved was manufactured in 2015 and the review of complaints revealed no other similar occurrences involving this unit.Thus, the event is an isolated case.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 ROLLER PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80939
GM   80939
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key17117862
MDR Text Key317059775
Report Number9611109-2023-00259
Device Sequence Number1
Product Code DTQ
UDI-Device Identifier04033817900382
UDI-Public010403381790038211151013
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K210130
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date 05/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/13/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number10-80-00
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received08/03/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/13/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-