Model Number PM2152
Device Problems Premature Discharge of Battery (1057); Failure to Interrogate (1332); No Pacing (3268)
Patient Problems Arrhythmia (1721); Dyspnea (1816)
Event Date 05/16/2023
Event Type  Injury  
Event Description
It was reported, the patient presented in the emergency room due to dyspnea and bradycardia.The device was unable to be interrogated and there was no magnet response.Premature battery depletion was suspected.The device was explanted and replaced to resolve the event.The patient was stable.
 
Manufacturer Narrative
The reported events of premature discharge of battery, no magnet response and no lv output were confirmed.The device was received with no telemetry and no output.Visual inspection of the header attachment area detected an anomaly between the pre-cast header and titanium case.The device was cut open to enable further testing and the battery was found below end-of-service voltage level.Feedthrough leak test was performed, indicating a device hermeticity breach.This is consistent with feedthrough damage as a result of fluid intrusion between the header and case, and subsequent fluid ingress to internal electronics.Hybrid circuitry was tested, and the results indicated elevated current drain, consistent with moisture damage, resulting in the reported events.A manufacturing process anomaly may have occurred, which resulted in the header bonding anomaly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDURITY¿ CORE DR, JPD
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17118021
MDR Text Key317090201
Report Number2017865-2023-23051
Device Sequence Number1
Product Code NVZ
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P880086
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 10/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2021
Device Model NumberPM2152
Device Lot NumberP000101171
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/31/2023
Initial Date FDA Received06/13/2023
Supplement Dates Manufacturer Received06/14/2023
10/13/2023
Supplement Dates FDA Received06/15/2023
10/19/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/17/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-